NAGCO
[NEW] Butadiene (C₄H₆) Polymer Detector Tube | Range: 0.1–200 ppmv | N012-3024-010
[NEW] Butadiene (C₄H₆) Polymer Detector Tube | Range: 0.1–200 ppmv | N012-3024-010
Designed for the measurement of Butadiene concentrations using a 9.8 eV standard lamp.
• Compound: Butadiene (C₄H₆)
• Typical Range: 0.1–200 ppmv
• Standard Lamp: 9.8 eV
• Sample Time: 75 seconds
• Sample Volume: 550 mL
• Total VOC Capacity: 150 ppmv
• Temperature Range: 5–40°C (41–104°F)
• Humidity Range: 0–95% RH
• Color Change: None
Storage: Unopened tubes can be stored for up to 2 years in darkness at 0–40°C (32–104°F). Opened tubes may be stored for up to 8 hours in clean air without significant loss of capacity.
Calibration: Calibration should be performed at the same temperature as the measurement. When calibrated from a dry gas cylinder, multiply the reading by the appropriate correction factor to obtain the true value.
Caution: Dispose of spent or expired tubes according to local regulations. Used tubes may contain toxic chemicals absorbed from the test environment. Unused tubes contain no hazardous or leachable materials.
Couldn't load pickup availability